1. Installing Your Steri-Trac™ Sensor 1 Installing Your Steri-Trac™ New Sensor Step One: Connect the sensor module to the monitor. The modul e slides on as shown in Figure 4. Be sure to squeeze the release tab while sliding the module onto the monitor as shown in Figure 5. Step Two: Plug in the monitor to electrical power. When power ed, the monitor will do the following: • All the LED’s in the display will illuminate red, t hen green. • The monitor will then display its firmware version number. • Then the monitor will display READING. • It will then display its initial ppm (parts per mil lion) reading. Step Three: Wait at least two minutes before perfor ming this next step. Then perform the zero-adjust function as described on th e next page using the “Magic Button™” (See Figure 6 below) as described on the n ext page. Figure 4: Slide the Sensor Module onto the monitor as shown. Figure 5: Squeeze the release tab (on the back of the sensor module) as shown when you connect the module to the monitor. Figure 6 Magic Button™/ Alarm silence button

3. Installing Your Steri-Trac™ Sensor 3 Step Four: Set Your Alarm Levels Your Steri-Trac Monitor has configurable alarm leve ls. When the monitor is shipped to you, its default low alarm level is 5ppm. The defau lt high alarm level is 45 ppm. If the monitor detects the target gas has reached 5ppm, th e low alarm will sound (a pulsating alarm). At 45 ppm the high alarm will sound (a cont inual alarm). To change alarm settings, do the following: Note: Once you enter the configuration menu, if the monitor detects no button presses for five seconds, it will exit the menu and return to t he main display. 1. To enter the configuration menu, press and hold the Magic Button™ until you see MENU on the main display. Then release the button. 2. Press and hold the button to cycle through the sub menus. The dis play will repeat cycling through the sub-menus as long as you hold d own the button. When you see LALM [low alarm] displayed, release the button. 3. Press and hold the button. The alarm setting will increase in val ue. Release the button when you see the value you want. (To alterna te between increasing or decreasing the setting, release the button and then press and hold again .) 4. If the monitor detects no button presses for five s econds, the monitor will exit the configuration menu, and then it will display in seq uence the low alarm setting, high alarm setting, and the horn volume. The monito r will then return to the main display. Your alarm level adjustment has been saved . To change the high alarm setting, repeat the above steps, but in step two enter the HALM menu instead LALM. End of Instructions. Congratulations, You Are Finished!! If you would like to hear the alarm horn, follow th e instructions for setting the horn volume found on p. 17 of the Steri-Trac Monitor Model 1000 Operation Manual . For more information about your new Steri-Trac Moni tor, please refer to the operation manual.

2. Installing Your Steri-Trac™ Sensor 2 Zero Adjust Function: The zero-adjust function is performed after a calib rated sensor has been newly installed. Once the sensor has settl ed, and there is no target gas in the area, the area monitor should read 0.0ppm. If the readin g is slightly off, use the zero- adjust function to inform the monitor that it shoul d be reading 0.0ppm. This filters out minor variations caused by environmental factors. Important Warnings Regarding the Zero-Adjust Functi on: • The zero-adjust function is intended to be used onl y after a calibrated sensor has been newly installed. Using the zero-adjust function for any other purpos e could seriously degrade the accuracy of the area mo nitor’s readings . • The zero-adjust function should only be performed w hen there is no target gas present in the area. Do not perform the zero adjust functi on, for example, while sterilizers are in operation. • After a calibrated sensor has been installed, wait two minutes before performing the zero-adjust function. This allows ti me for the sensor to settle. Note: Once you enter the configuration menu, if the monitor detects no button presses for five seconds, it will exit the menu and return to t he main display. 1. To enter the configuration menu, press and hold the Magic Button™ until you see MENU on the main display. Then release the button. 2. Press and hold the button to cycle through the sub menus. The dis play will repeat cycling through the sub menus as long as you hold d own the button. When you see ZERO displayed, release the button. 3. With ZERO displayed, press the button once and release. The monitor will go through its zero-adjust function and will flash in green Z0.0. 4. After a few seconds, the monitor will display in se quence the low alarm setting, high alarm setting, the horn volume, then it will d isplay Z0.0. If the Z0.0 is GREEN, the zero adjust was successful. (If the Z0.0 is RED, the zero-adjust was NOT successful. Call your ChemDAQ service representative.) 5. The monitor will then return to the main display.

Views

  • 98 Total Views
  • 67 Website Views
  • 31 Embedded Views

Actions

  • 0 Social Shares
  • 0 Likes
  • 0 Dislikes
  • 0 Comments

Share count

  • 0 Facebook
  • 0 Twitter
  • 0 LinkedIn
  • 0 Google+